LRQC (802)    OPTION (19)

View in ViViaN Menu
Name Value
NAME LRQC
MENU TEXT Quality control display (Levey-Jennings)
UPPERCASE MENU TEXT QUALITY CONTROL DISPLAY (LEVEY
PACKAGE LAB SERVICE
ROUTINE LRQC
DESCRIPTION
Quality control data are displayed against the normal mean and standard
report also has a line where the responsible laboratory official 
may review the output of the report (CAP Requirement).  The
"mean" is denoted by the ***** line, 1SD & 3SD are denoted by the 
..... lines, and 2SD is denoted by the ::::: lines.
deviation entered for the requested test (in the Test field of the
Lab Control Name file).  The option uses a modified Levey-Jennings
format to allow a printout on the VA printers.  This format is 
vertical and the time scale is sequential and not proportional.
The report will list all dates, control values, total # of
controls (N), target range, actual range obtained, and will flag
any values outside of 3SD.  The control values outside of 3SD are 
not used in the calculation of the actual range obtained.  The
CREATOR POSTMASTER
TYPE run routine