Name | Value |
---|---|
NAME | VITEK |
METHOD | VITEK |
DEFAULT ACCESSION AREA | MICROBIOLOGY |
OVERLAY DATA | YES |
PROGRAM | LAMIVTK |
NEW DATA | D NEW^LASET |
RESTART | D RESTART^LASET |
HANDSHAKE RESPONSE | S OUT=$C(6) |
LOAD/WORK LIST | File: 68.2, IEN: 15 |
ENTRY for LAGEN ROUTINE | Accession cross-reference |
CROSS LINKED BY | +ID |
MICRO CARD TYPE |
|
INTERFACE NOTES | The electrical connection of the VITEK can be done with the following. 20--| up the Autoinstrument node (62.4). See lab installation guide for procedures on adding a new organism to the etiology file. STEP 3. Step thru the beginning fields and fill them in as you would normally do. Several fields will need to be set up differently depending on whether you are running unidirectionally or bidirectionally. These fields are: Program Name uni = LAMIVTK bi = LAMIVTKU Handshake response uni = S OUT=$C(6) bi = D ^LAMIVTKC STEP 4. The entry comes with 10 card types already started for you. These entries are never to be deleted as other entries will refer back to them. Using your VITEK interface documentation enter the card type you will want to enter organisms or drugs for. Step thru the prompts until you come to the Select Organism prompt. Enter the organism name and its card code for each organism for this card type. Once you have organisms entered for the intitial 10 entries you do not have to enter organisms for any of the following card types as the code will refer back to these 7------------------------7 initial entries. When entering a new card the process card call field needs to be filled in with the MUMPS code listed below for PRocessing call. For example if you want to process a GNS-F1 card which has a card code of 54 you would enter 'D ^LAMIV12'. Card Ref. Code Spec. Name Processing call 01 5.1 GPS-A D ^LAMIV11 02 5.3 GPSSA D ^LAMIV11 03 5.5 GPSTA D ^LAMIV11 04 NOT USED 05 5.4 YBC D 54^LAMIV11 06-0A CARD TYPES NOT USED 0B 5.10 UID-1 D ^LAMIV10 0C 5.2 BACIL D 52^LAMIV10 0D XXX BIOLD NOT SUPPORTED BY VITEK 0E 5.12 GNI D 512^LAMIV12 0F 5.11 GPI D 511^LAMIV11 To set up the VITEK unidirectionally you will need to run the program 0FB 5.11 GPI-2 Mate to '0F' which holds 2nd organism table. Called from '0F' card if needed. 10 5.13 GPS-M D ^LAMIV11 11 XXX SCC NOT SUPPORTED BY VITEK 12 5.10 UID-3 D 510^LAMIV10 13 5.14 GNS D ^LAMIV12 14 5.15 GNU D ^LAMIV12 15 5.16 GNS-T D ^LAMIV12 16 5.17 GNU-T D ^LAMIV12 17-A4 GNS FLEX D ^LAMIV12 CIEP on the VITEK to configure the instrument. If you are running with Since so many of the cards have the same organisms identified they are entered only once on one type of card. The Gram Positive cards use 1 of 2 tables of organisms which are listed under card types '0F' and '0FB' and these tables are used for lookup of the organism for all gram positive identifications whether it is on an GPI card or on a GPS flex card. The Gram Negative codes are handled by a single table identified under the GNI card and is used for all flex panals also. Gram Negative flex cards can list up to 5 organisms on one card and when this occurs the organism list is found on the GNS card. bidirectional you will need to run the program BCIEP to accomplish this. The following is a list of where the organisms will be found for the different cards. Card '0F' is the main GPI card, and is used by all GPS cards for Org ID lookup. Card '0FB' is the secondary GPI card used to hold second table of organisms. It is used by processing routine when needed. This card is a Pseudo card and therefore does not have a call. Card '0E' is the main GNI card, and is used by all GNS cards for Org ID lookup. CARD '10' is the main GPS flex card and is used by all other flex cards for Organism lookup when multiple organisms are identified. CARD '05' is the main card used for identifying yeasts. CARD '0B' is used to identify urines CARd '12' is also used to identify urines. CARD '13' is the GNS flex card and is used by all other flex cards for organism lookup when multiple organisms are identified. CARD '14' is the GNU flex card and is used by all other flex cards for organism lookup when multiple organisms are identified. To set up the VITEK for unidirectional the following paramaters should CARD '0C' is used for Bacillus identification. STEP 5 Now step to the Select drug prompt for the card type. If the documentation has a list of drugs for that card type then enter the drug name, its param 1, its card code, and the print order. Many of the drugs listed for a card type have more than one code associated with the drug. When this occurs you will add the drug name to the listing once for each code associated with that drug. be set as: The VITEK data is processed by a set of routines that start LAMIV. LAMIV00 is routine used to determine MIC value for the drug. The routine is called from the param1 field of the drug subfield of the micro card type field. The param1 is in the form: "S RUN="x"_U_LARTN D @RUN where "x" is the line tag to enter the routine LAMIV00.. The Tag line in the routine LAMIV00 for each entry point tells which drug codes and type of card (pos, neg) are valid for this entry point. For example if you are using a gram negative card which VITEK LSI reports out the drug code as 'C' for tetracycline you would find by looking at the routine LAMIV00 that the proper entry point would be at 'I' and the entry in param1 would then be 'S LARTN="I"_U_"LAMIV00" D @LARTN'. LAMIV10 is routine used for processing bacillus and urine ID cards LAMIV11 is routine for processing Gram Positive identification, flex cards and Yeast cards. LAMIV12 is routine for processing Gram Negative identification and Gram Negative flex cards cards. ON/OFF SWITCH: OFF The above routines will convert the VITEK coded data to an organism name, a drug name, a raw MIC value, and an interpretation which will all be shown to you upon verification. Certain drugs have codes which can not be entered thru fileman. These drug codes are "^@?". These characters have special meaning to FILEMAN and cannot be entered directly. To get around this problem the routine LAMICRA is included. This routine will prompt for the autoinstrument entry, the card type and then walk thru the drug nodes displaying the entered code. You can enter the code or enter a return at this time. RETRY WAIT: 10 SECONDS (SET TO MAX ALLOWED) The routine will either continue when a return is entered or enter the code into the file and build the "C" cross reference entry for you. ****NOTE**** If the autoinstrument file is ever recross referenced you will have to reenter these codes as they will be lost. The last field here is the print order field. This field is used by the verify code to display the list of drugs in whatever order you chose. This was primarily added so that the organisms and drugs would display in the same order as the printed VITEK form. ACK WAIT: 3 SECONDS (SET TO MAX ALLOWED) Occasionally VITEK will start producing a new card type. These cards will usually have a card code above A4. If you want to use the new card be sure of the type of card it is and use the process card call that is used for a similar card. When running in bidirectional mode some additional fields need to be filled in. The FILE BUILD ROUTINE field needs to have the program name LAMIVTKD NACKS BEFORE AUTO DISABLE: 3 which builds and stores the download records from the load list file. In the Lab Site Parameter file you will need to respond to the new field DOWNLOAD FULL DATA. If you answer YES the routines will download all of the data available about the patient and test. If you answer NO the routines will only download the minimum information needed by the instrument to run the test. In the LAB TEST file those tests that are to be downloaded will need a unique number assigned to it (0-9) in the field MICRO PREFIX CODE. STX/ETX PROTOCOL: "SSS"/<CR><LF> The auto instrument file has 3 additional fields to fill in. Send tray/cup - which is a yes/no answer and is the default display when you start generating the download file. Queue build - is a yes/no answer that is the default display when starting to generate a download file and determines if the file is built while you wait or is queued out to run by itself giving you you terminal back immediately. Be very aware that the verify routines used by the rest of the LAB will HOST PROTOCAL CHARACTERISTICS: 1200 BAUD, NO PARITY, 8 BIT/CHAR, NOT work for microbiology. Use only the verify routines for micro which will be supplied. 2 STOP BITS XON TIME-OUT VALUE: 10 DELAY BETWEEN CHARACTERS: 50 2---------\/-------------2 DISPLAY DATA OPTION: OFF AMS COMPUTER INTERFACE: ENABLED For bidirectional you have the following parameters. BIDIRECTIONAL INTERFACE (UPLOAD) IS : ENABLED BIDIRECTIONAL INTERFACE (DOWNLOAD) IS: ENABLED DOWNLOAD PORT: /dev/tty1 UPLOAD PORT: /dev/tty1 TIMEOUTS: CHECKSUM <ACK>: default is 3 set as high as possible 3---------/\-------------3 HOST RESPONSE: defalut is 3 set as high as possible XON/XOFF: default is 10 set higher if having problems RETRY LIMITS: CHECKSUM ERROR: 3 <ENQ>: 3 RETRY INTERVALS: CHECKSUM ERROR: 10 <ENQ>: 10 DELAYS: LAST MASTER: DEFAULT 10 SET HIGHER 4--| INTER-RECORD: DEFAULT 10 SET HIGHER INTERMESSAGE: DEFAULT 10 SET HIGHER END OF RECORD FORMATS STX: ENABLED ETX: DISABLED RS: DISABLED GS: DISABLED TOTAL NUMBER OF FAILED MESSAGED ALLOWED: 3 FIELD TERMINATION CHARACTER(S): | START UP MESSAGE: DISABLED In some instances the VITEK will not recieve our acknowledge back in the 5--| 10 second time limit of VITEK and will retransmit the record a second time. If this is occuring from the console of the VITEK enter /ETC/HOSTCTS OFF This command will in most cases clear up double sending of records. The VITEK has one caution that must be remembered. If during transmission the VITEK does not receive acknowledgement from the CPU that the records had been received within 3 tries it will turn the interface off. When this occurs several things must be remembered. 1. The VITEK will not transmit again until the operator turns the interface back on. 2. Once the interface has been turned back on the operator can retransmit records ONLY if the cards have NOT been removed from the machine. If they have been removed the entire test must be rerun. This node of the Auto Instrument file is configured to support revision AI of the VITEK software. Sometime around release of Revision L there was a change in the reporting of values out of the VITEK. Then with the release of Revisions W and above to Revision AI they have periodically changed values being reported. The entries in this file have been set up for Revision AI and therefore NOTE that we will support only revision AI or 6--| later. The VITEK has several tables which need to be built before the instrument can be run. These tables have an entry and then a abbrreviation. These tables are set up to match your DHCP tables. Now we are ready to start setting up the VITEK entry which has several steps that must be followed in sequence in order to set the entry up properly. 8--|---------------------20 STEP 1. The VITEK uses a large number of drugs when doing its reporting. You MUST have all drugs entered into the Antimicrobial Sensitivity file (62.06) and the Lab Data File (63) before setting up the Autoinstrument file (62.4). See the Lab installation guide for particulars for adding new drugs to the system. STEP 2. Also since the VITEK can report over 100 different organisms make sure that all the organisms are entered into the Etiology file (61.2) before setting |
*ECHO ALL INPUT | NO |
FILE BUILD ROUTINE | LAMIVTKD |
SEND TRAY/CUP LOCATION | no |
QUEUE BUILD | no |