Name | Value |
---|---|
NAME | EKTACHEM 700 BI |
METHOD | EKTCH700 |
DEFAULT ACCESSION AREA | CHEMISTRY |
OVERLAY DATA | YES |
PROGRAM | LAEKT7B |
NEW DATA | D NEW^LASET |
HANDSHAKE RESPONSE | D ^LAKERMIT |
ACK TRIGGER VALUE | 0 |
ACK RESPONSE VALUE | 0 |
LOAD/WORK LIST | File: 68.2, IEN: 8 |
LABORATORY TESTS |
|
LOAD CHEM TESTS | TMP GLOBAL |
ENTRY for LAGEN ROUTINE | Accession cross-reference |
CROSS LINKED BY | IDE |
INTERFACE NOTES | LAEKT7B is used to process output taken from the instrument interface. will be in the 3rd piece. When the EKTACHEM 700 is setup to run in the BI-directional mode, and the site has the new LSI rom. The best setings to use are 7 or 8 bits with ODD parity and 1 stop bit. This routine can't be used in the Uni-directional mode. The routine calls LAKERM3 as part of the IN subroutine to unpack the raw kermit data from ^LA(tsk,"O",n) to ^LA(tsk,"C",n), then parses the tests into TY(testcode)=data^codes. If the error flag has a value other than 0,1,2,7 then the value will be <null> and the tranmited value |
DOWNLOAD ENTRY | DATA |
DOWNLOAD PROTOCOL ROUTINE | LAKERM2 |
FILE BUILD ROUTINE | LAEKT7D |