DIRECT EPX BI (186)    AUTO INSTRUMENT (62.4)

Name Value
NAME DIRECT EPX BI
METHOD EPX
DEFAULT ACCESSION AREA CHEMISTRY
OVERLAY DATA YES
PROGRAM LAEPX
NEW DATA D NEW^LASET
RESTART D RESTART^LASET
LABORATORY TESTS
  • TEST:   CO2
    ROUTINE STORAGE:   TV(8,1)
    PARAM 3:   BICAR
    UI TEST CODE:   BICAR
  • TEST:   CALCIUM
    ROUTINE STORAGE:   TV(9,1)
    PARAM 3:   CALCIUM
    UI TEST CODE:   CALCIUM
  • TEST:   CHOLESTEROL
    ROUTINE STORAGE:   TV(12,1)
    PARAM 3:   CHOL
    UI TEST CODE:   CHOL
  • TEST:   CPK
    ROUTINE STORAGE:   TV(41,1)
    PARAM 3:   CK
    UI TEST CODE:   CK
  • TEST:   DIR. BILIRUBIN
    ROUTINE STORAGE:   TV(16,1)
    PARAM 3:   DBILI
    UI TEST CODE:   D BILI
  • TEST:   GAMMA-GTP
    ROUTINE STORAGE:   TV(21,1)
    PARAM 3:   GGTP
    UI TEST CODE:   GGTP
  • TEST:   ALBUMIN
    ROUTINE STORAGE:   TV(14,1)
    PARAM 3:   ALBUMIN
    UI TEST CODE:   ALBUMIN
  • TEST:   GLOBULIN
    ROUTINE STORAGE:   TV(260,1)
    PARAM 3:   GLOBULIN
    UI TEST CODE:   GLOBULIN
  • TEST:   GLUCOSE1
    ROUTINE STORAGE:   TV(2,1)
    PARAM 3:   GLUCOSE
    UI TEST CODE:   GLUCOSE
  • TEST:   HDL
    ROUTINE STORAGE:   TV(80,1)
    PARAM 3:   HDL
    UI TEST CODE:   HDL
  • TEST:   POTASSIUM
    ROUTINE STORAGE:   TV(6,1)
    PARAM 3:   K
    UI TEST CODE:   K
  • TEST:   LDH1
    ROUTINE STORAGE:   TV(131,1)
    PARAM 3:   LD1
    UI TEST CODE:   LD1
  • TEST:   MAGNESIUM
    ROUTINE STORAGE:   TV(29,1)
    PARAM 3:   MAG
    UI TEST CODE:   MAG
  • TEST:   SODIUM
    ROUTINE STORAGE:   TV(5,1)
    PARAM 3:   NA
    UI TEST CODE:   NA
  • TEST:   OSMOLARITY
    ROUTINE STORAGE:   TV(46,1)
    PARAM 3:   OSMOL
    UI TEST CODE:   OSMOL
  • TEST:   PO4
    ROUTINE STORAGE:   TV(10,1)
    PARAM 3:   PHOS
    UI TEST CODE:   PHOS
  • TEST:   TOT. BILIRUBIN
    ROUTINE STORAGE:   TV(15,1)
    PARAM 3:   TBILI
    UI TEST CODE:   T BILI
  • TEST:   PROTEIN,TOTAL
    ROUTINE STORAGE:   TV(13,1)
    PARAM 3:   TPROT
    UI TEST CODE:   T PROT
  • TEST:   UREA NITROGEN
    ROUTINE STORAGE:   TV(3,1)
    PARAM 3:   UREA
    UI TEST CODE:   UREA
  • TEST:   URIC ACID
    ROUTINE STORAGE:   TV(11,1)
    PARAM 3:   URIC
    UI TEST CODE:   URIC
  • TEST:   AMMONIA
    ROUTINE STORAGE:   TV(48,1)
    PARAM 3:   AMM
    UI TEST CODE:   AMM
  • TEST:   AMYLASE
    ROUTINE STORAGE:   TV(40,1)
    PARAM 3:   AMY
    UI TEST CODE:   AMY
  • TEST:   ANION GAP
    ROUTINE STORAGE:   TV(790,1)
    PARAM 3:   ANION
    UI TEST CODE:   ANION
LOAD CHEM TESTS TC ARRAY
ENTRY for LAGEN ROUTINE Accession cross-reference
CROSS LINKED BY ID
INTERFACE NOTES
 Re Direct Connect (bypassing interface) :
          CRT                 : Y
          Login Allowed       : Y
          Output Margin       : 255
          Tie Rtn             : 1 (optional)
          Auto Baud           : N
          ZUSE                : Y
 
     RS 232 : same as terminal; Pins 2, 3, and 7 with Pin 4 connected to 
                                Pin 5 and Pin 6 connected to Pin 20. 
                                ( Latter 2 connects short modem contrl )
 
 
     EPX : Serial Out @ 1200 baud, 8 Data, 1 Stop, No Parity.
 
Re Setting files for operation:
  AUTO INSTRUMENT FILE:
     When entering tests into the CHEM TESTS field be sure the PARAM 3
     subfields have no spaces in the test names entered here.
  LOAD/WORKLIST FILE:
     In order to display the tests for verification using EA in the
     same order as they come off on the EPX printouts you must:
     MUX settings:
     1)  Create a test in the LAB TEST FILE (e.g. EPX PROFILE) with
         the tests in LAB TESTS INCLUDED IN PANEL entered in the order
         they display from your EPX printout.  TYPE is OUTPUT.
     2)  Enter this test into the LOAD/WORKLIST Profile you have
         given for the EPX.  This test should be BUILD NAME ONLY: NO.
         All the other tests should be set to BUILD NAME ONLY: YES.
     3)  LOAD TRANSFORM should be set to UNIVERSAL.  INITIAL SETUP
         needs to be set to INITIAL SET.
From ABBOTT
     In regard to smooth operation of the Downloading routine:
          Speed (RCVR & XMT)  : 1200
     1)  Must reindex the disk at least weekly, if not more frequently.
     2)  Use a separate carousel (e.g. 99) for running controls,
         calibrators, and water.  If system prompts for these
         parameters should appear during a run on your test
         carousel they will interfere with the download.
     3)  You may use the same carousel until it is full.  The routines
         will progress to the next carousel.  Do not use a carousel
         that has previously been employed without first doing the
         reindex (see #1 above).
     4)  Auto send- ON
          Modem Cntrl         : NONE
     5)  The rationale for the reindexing is that current EPX software
         will not consistently delete the PID which is assigned during
         the download.  This causes problems when trying subsequent
         downloads to the same carousel.  Hence, reindex or use
         a different carousel.
     6)  Everytime there is a glitch with the host computer the
         EPX stops talking to the host and must be rebooted.
          Stall Count         : 0
          Output Only         : N
          Tab Cntrl           : Y
          Lowcase Cntrl       : Y
*ECHO ALL INPUT NO
FILE BUILD ROUTINE LAEPXD