INTERFACE NOTES |
Sample LA format:
^LA(5,"I",5) = R|853|63 |003|1|1| 28.2|0| |0| |000|0| |0|0| |
PARAM 3: should have the test identifier entered for all tests except
the INR. The identifier number can be found in the Test ID section of
the MLA documentation. The most common ones are: PT=1, APTT=3,
Fibrinogen=4, TT=5. The documentation will list tehm as 001 or 012 but you
should just enter the number without the preceding zeros.
PARAM 1 is used to hold the extract code for the INR, it should be:
S V=+Y(2) D NUM
Wiring requirements are:
^LA(5,"ENV") = VAH^ROU^50000VA0
BAUD RATE=1200, PARITY=none, STOP BITS=1, CHARACTER BITS=8 (standard
unmodified LSI port). Set instrument to SEND DURING RUN=ON,
DATA FORMAT=E1000C. Pins 2 and 3 will need to be reversed.
^LA(5,"I") = 6
^LA(5,"I",O) = 6
^LA(5,"I",1) = R|848|61 |001|1|1| 13.9|0| |0| |000|0| |0|0|1.2|
^LA(5,"I",2) = R|849|61 |003|1|1| 42.5|0| |0| |000|0| |0|0| |
^LA(5,"I",3) = R|850|62 |001|1|1| 12.3|0| |0| |000|0| |0|0|0.9|
^LA(5,"I",4) = R|851|62 |003|1|1| 27.7|0| |0| |000|0| |0|0| |
^LA(5,"I",5) = R|852|63 |001|1|1| 12.0|0| |0| |000|0| |0|0| |
|