MLA1000C (193)    AUTO INSTRUMENT (62.4)

Name Value
NAME MLA1000C
METHOD MLA1000C
DEFAULT ACCESSION AREA COAG
OVERLAY DATA YES
PROGRAM LAMLA1KC
NEW DATA D NEW^LASET
RESTART D RESTART^LASET
LOAD/WORK LIST File: 68.2, IEN: 21
LABORATORY TESTS
  • ROUTINE STORAGE:   TV(430,1)
ENTRY for LAGEN ROUTINE Accession cross-reference
CROSS LINKED BY ID
INTERFACE NOTES
Sample LA format:
^LA(5,"I",5) = R|853|63  |003|1|1| 28.2|0|  |0|  |000|0| |0|0|   |
 
PARAM 3: should have the test identifier entered for all tests except 
the INR.  The identifier number can be found in the Test ID section of 
the MLA documentation.  The most common ones are: PT=1, APTT=3, 
Fibrinogen=4, TT=5. The documentation will list tehm as 001 or 012 but you 
should just enter the number without the preceding zeros.
PARAM 1 is used to hold the extract code for the INR, it should be: 
S V=+Y(2) D NUM
Wiring requirements are:
^LA(5,"ENV") = VAH^ROU^50000VA0
BAUD RATE=1200, PARITY=none, STOP BITS=1, CHARACTER BITS=8 (standard
unmodified LSI port). Set instrument to SEND DURING RUN=ON, 
DATA FORMAT=E1000C. Pins 2 and 3 will need to be reversed.
^LA(5,"I") = 6
^LA(5,"I",O) = 6
^LA(5,"I",1) = R|848|61  |001|1|1| 13.9|0|  |0|  |000|0| |0|0|1.2|
^LA(5,"I",2) = R|849|61  |003|1|1| 42.5|0|  |0|  |000|0| |0|0|   |
^LA(5,"I",3) = R|850|62  |001|1|1| 12.3|0|  |0|  |000|0| |0|0|0.9|
^LA(5,"I",4) = R|851|62  |003|1|1| 27.7|0|  |0|  |000|0| |0|0|   |
^LA(5,"I",5) = R|852|63  |001|1|1| 12.0|0|  |0|  |000|0| |0|0|   |