INTERFACE NOTES |
To enable the interface, press SYSTEM, then FILES, enter 1.18, then press
crossed so that it will transmit as a DTE interface.
^LA(19,"I",26) = --------------------------------
^LA(19,"I",27) = A 0. 6.2
^LA(19,"I",28) = B 2. 71.4
^LA(19,"I",29) = C 10. 294.6
^LA(19,"I",30) = D 30. 720.2
^LA(19,"I",31) = E 60. 1157.2
^LA(19,"I",32) = F 100. 1539.7
^LA(19,"I",33) =
^LA(19,"I",34) = LOC ID ng/mL RATE NOTE
^LA(19,"I",35) = ----------------------------------------
^LA(19,"I",36) = 2 CONTROL L 3.88 128.0
^LA(19,"I",37) = 3 CONTROL M 14.17 396.3
^LA(19,"I",38) = 4 CONTROL H 42.42 923.2
^LA(19,"I",39) = 5 000000400 0.68 29.5 LOW
^LA(19,"I",40) = 6 000000191 1.24 47.5 LOW
^LA(19,"I",41) = 7 000000301 2.54 88.0
^LA(19,"I",42) =
^LA(19,"I",43) =
^LA(19,"I",44) = HIGH > 100.000 ng/mL
^LA(19,"I",45) = LOW < 2.000 ng/mL
The LAIMX routine is written to process data that is transmitted in
^LA(19,"I",46) =
^LA(19,"I",47) =
^LA(19,"I",48) = TESTS USED TO DATE : 61
^LA(19,"I",49) =
^LA(19,"I",50) =
^LA(19,"I",51) =
^LA(19,"I",52) =
^LA(19,"I",53) = ]
^LA(19,"I",54) = C
^LA(19,"I",55) = !!
the MEIA Quantitative layout as defined in the IMX unidirectional RS232
Interface Guide, pages 7-10.
Param 2 holds the number of decimal places for the result.
Param 3 holds the number of the test as identified by the instrument.
This interface will transmit in a multiple of different formats depending
DISPLAY. Enter the password 2215 and press STORE. Set the
on the tests being run. This version as of 8/13/93 will handle the data
streams for the ASSAY NUMBER, %GHb and %A1C tests, as well as linearity
checking.
**The %A1C test will require the following code to be placed into Param
1: S LATFLG=1. All other tests should not have this variable set. The
linearity functionality can be controlled by entering the following code
in Param 1: S LRLN(III)="X.X" where the X.X represents the linearity value
for this test. If no linearity checking is to be enabled, Param 1 should
be set LRLN(III)="0.0".
communication defaults for the COM2 port, turn the IMX system
Select TEST: PSA
ARE YOU ADDING 'PSA' AS A NEW CHEM TESTS (THE 1ST FOR THIS AUTO
INSTRUMENT)? Y (YES)
CHEM TESTS NUMBER: 1//
PARAM 1: S LRLN(III)="0.1"
PARAM 2: 1
PARAM 3: 53
DOWNLOAD CODE:
Select TEST: CEA
ARE YOU ADDING 'CEA' AS A NEW CHEM TESTS (THE 2ND FOR THIS AUTO
off, wait 10 seconds, then turn it back on. This initializes the
INSTRUMENT)? Y (YES)
CHEM TESTS NUMBER: 2//
PARAM 1: S LRLN(III)="1.0"
PARAM 2: 1
PARAM 3: 76
DOWNLOAD CODE:
Select TEST: FERRITIN
ARE YOU ADDING 'FERRITIN' AS A NEW CHEM TESTS (THE 3RD FOR THIS AUTO
INSTRUMENT)? Y (YES)
CHEM TESTS NUMBER: 3//
parameters in memory.
PARAM 1: S LRLN(III)="1.0"
PARAM 2: 0
PARAM 3: 27
DOWNLOAD CODE:
Select TEST: HEMOGLOBIN A1C
ARE YOU ADDING '"HEMOGLOBIN A1C' AS A NEW CHEM TESTS (THE 4TH FOR THIS
AUTO INSTRUMENT)? Y (YES)
CHEM TESTS NUMBER: 4//
PARAM 1: S LRLN(III)="0.0"
PARAM 2: 1
PARAM 3: 79
DOWNLOAD CODE:
Select TEST: (continue entering all tests that this instrument will run
using PARAM 3 to hold the ASSAY # assigned by Abbott.)
******************************************************************
SAMPLE DATASTREAM
FYI: This instrument produces several different data streams not all of
which are currently supported.
The interface may be tested by reprinting the results from the last run.
******************************************************************
^LA(19,"ENV") = VAH^ROU^583VA0^ROU583VA0
^LA(19,"I") = 55
^LA(19,"I",0) = 55
^LA(19,"I",1) = [
^LA(19,"I",2) =
^LA(19,"I",3) =
^LA(19,"I",4) = DATE: 4/27/93
^LA(19,"I",5) = TIME: 13:45:30
^LA(19,"I",6) =
^LA(19,"I",7) = TECH ID : _________
^LA(19,"I",8) = RGNT LOT: _________
^LA(19,"I",9) = SERIAL #: 22642
^LA(19,"I",10) = CRSL ID : 2
^LA(19,"I",11) = CURVE : 1
^LA(19,"I",12) = CAL DATE: 4/02/93
^LA(19,"I",13) = CAL TIME: 10:37:52
^LA(19,"I",14) =
^LA(19,"I",15) = ASSAY 53 PSA 2245 Revision 1
***The instrument uses a DCE configuration. Lines 2 and 3 must be
^LA(19,"I",16) =
^LA(19,"I",17) = MODE 1 CALIBRATOR
^LA(19,"I",18) =
^LA(19,"I",19) = LOC ng/mL RATE FACTOR
^LA(19,"I",20) = ----------------------------
^LA(19,"I",21) = 1 10. 294.6 0.926
^LA(19,"I",22) =
^LA(19,"I",23) = MODE 1 CALIBRATION CURVE
^LA(19,"I",24) =
^LA(19,"I",25) = CALIB ng/mL AVGR
|