ABBOTT IMX UNI (197)    AUTO INSTRUMENT (62.4)

Name Value
NAME ABBOTT IMX UNI
METHOD ABBOTT IMX
OVERLAY DATA YES
PROGRAM LAIMX
NEW DATA D NEW^LASET
RESTART D RESTART^LASET
ENTRY for LAGEN ROUTINE Accession cross-reference
CROSS LINKED BY IDE
INTERFACE NOTES
 To enable the interface, press SYSTEM, then FILES, enter 1.18, then press
 crossed so that it will transmit as a DTE interface.
 ^LA(19,"I",26) = --------------------------------
 ^LA(19,"I",27) =   A           0.     6.2
 ^LA(19,"I",28) =   B           2.    71.4
 ^LA(19,"I",29) =   C          10.   294.6
 ^LA(19,"I",30) =   D          30.   720.2
 ^LA(19,"I",31) =   E          60.  1157.2
 ^LA(19,"I",32) =   F         100.  1539.7
 ^LA(19,"I",33) =
 ^LA(19,"I",34) = LOC    ID            ng/mL     RATE NOTE
 ^LA(19,"I",35) = ----------------------------------------
 
 ^LA(19,"I",36) =  2  CONTROL L         3.88    128.0
 ^LA(19,"I",37) =  3  CONTROL M        14.17    396.3
 ^LA(19,"I",38) =  4  CONTROL H        42.42    923.2
 ^LA(19,"I",39) =  5  000000400         0.68     29.5  LOW
 ^LA(19,"I",40) =  6  000000191         1.24     47.5  LOW
 ^LA(19,"I",41) =  7  000000301         2.54     88.0
 ^LA(19,"I",42) =
 ^LA(19,"I",43) =
 ^LA(19,"I",44) = HIGH >   100.000   ng/mL
 ^LA(19,"I",45) =  LOW <     2.000   ng/mL
 The LAIMX routine is written to process data that is transmitted in
 ^LA(19,"I",46) =
 ^LA(19,"I",47) =
 ^LA(19,"I",48) = TESTS USED TO DATE :  61
 ^LA(19,"I",49) =
 ^LA(19,"I",50) =
 ^LA(19,"I",51) =
 ^LA(19,"I",52) =
 ^LA(19,"I",53) = ]
 ^LA(19,"I",54) = C
 ^LA(19,"I",55) = !!
 the MEIA Quantitative layout as defined in the IMX unidirectional RS232
 Interface Guide, pages 7-10.
 
 Param 2 holds the number of decimal places for the result.
 Param 3 holds the number of the test as identified by the instrument.

 This interface will transmit in a multiple of different formats depending
 DISPLAY. Enter the password 2215 and press STORE. Set the
 on the tests being run. This version as of 8/13/93 will handle the data
 streams for the ASSAY NUMBER, %GHb and %A1C tests, as well as linearity
 checking.
 **The %A1C test will require the following code to be placed into Param
 1: S LATFLG=1. All other tests should not have this variable set. The
 linearity functionality can be controlled by entering the following code
 in Param 1: S LRLN(III)="X.X" where the X.X represents the linearity value
 for this test. If no linearity checking is to be enabled, Param 1 should
 be set LRLN(III)="0.0".

 communication defaults for the COM2 port, turn the IMX system
 Select TEST: PSA 
  ARE YOU ADDING 'PSA' AS A NEW CHEM TESTS (THE 1ST FOR THIS AUTO 
          INSTRUMENT)? Y   (YES) 
    CHEM TESTS NUMBER: 1// 
  PARAM 1: S LRLN(III)="0.1" 
  PARAM 2: 1 
  PARAM 3: 53 
  DOWNLOAD CODE: 
 Select TEST: CEA 
  ARE YOU ADDING 'CEA' AS A NEW CHEM TESTS (THE 2ND FOR THIS AUTO 
 off, wait 10 seconds, then turn it back on. This initializes the
         INSTRUMENT)? Y  (YES) 
     CHEM TESTS NUMBER: 2// 
  PARAM 1: S LRLN(III)="1.0" 
  PARAM 2: 1 
  PARAM 3: 76 
  DOWNLOAD CODE: 
 Select TEST: FERRITIN 
   ARE YOU ADDING 'FERRITIN' AS A NEW CHEM TESTS (THE 3RD FOR THIS AUTO 
         INSTRUMENT)? Y  (YES) 
     CHEM TESTS NUMBER: 3// 
 parameters in memory.
  PARAM 1: S LRLN(III)="1.0" 
  PARAM 2: 0 
  PARAM 3: 27 
  DOWNLOAD CODE: 
 Select TEST: HEMOGLOBIN A1C 
   ARE YOU ADDING '"HEMOGLOBIN A1C' AS A NEW CHEM TESTS (THE 4TH FOR THIS 
          AUTO INSTRUMENT)? Y  (YES) 
     CHEM TESTS NUMBER: 4// 
  PARAM 1: S LRLN(III)="0.0" 
  PARAM 2: 1 

  PARAM 3: 79 
  DOWNLOAD CODE: 
 Select TEST: (continue entering all tests that this instrument will run 
              using PARAM 3 to hold the ASSAY # assigned by Abbott.) 

 
 ******************************************************************
                SAMPLE DATASTREAM
 FYI: This instrument produces several different data streams not all of
 which are currently supported.
 The interface may be tested by reprinting the results from the last run.
 ******************************************************************
 
 ^LA(19,"ENV") = VAH^ROU^583VA0^ROU583VA0
 ^LA(19,"I") = 55
 ^LA(19,"I",0) = 55
 ^LA(19,"I",1) = [
 ^LA(19,"I",2) =
 ^LA(19,"I",3) =
 ^LA(19,"I",4) = DATE:  4/27/93
 ^LA(19,"I",5) = TIME: 13:45:30
 
 ^LA(19,"I",6) =
 ^LA(19,"I",7) = TECH ID : _________
 ^LA(19,"I",8) = RGNT LOT: _________
 ^LA(19,"I",9) = SERIAL #: 22642
 ^LA(19,"I",10) = CRSL ID :   2
 ^LA(19,"I",11) = CURVE   : 1
 ^LA(19,"I",12) = CAL DATE:  4/02/93
 ^LA(19,"I",13) = CAL TIME: 10:37:52
 ^LA(19,"I",14) =
 ^LA(19,"I",15) = ASSAY  53 PSA         2245 Revision 1
 ***The instrument uses a DCE configuration. Lines 2 and 3 must be
 ^LA(19,"I",16) =
 ^LA(19,"I",17) = MODE 1 CALIBRATOR
 ^LA(19,"I",18) =
 ^LA(19,"I",19) = LOC     ng/mL    RATE FACTOR
 ^LA(19,"I",20) = ----------------------------
 ^LA(19,"I",21) =  1        10.   294.6  0.926
 ^LA(19,"I",22) =
 ^LA(19,"I",23) = MODE 1 CALIBRATION CURVE
 ^LA(19,"I",24) =
 ^LA(19,"I",25) = CALIB      ng/mL    AVGR