| DESCRIPTION OF ENHANCEMENTS |
This patch addresses the following issue:
alter or modify any software design safeguards or safety critical
File Name Contents Retrieval Format
------------------ -------------------- ----------------
LAB_MICRO_INTERFACE_RELEASE_1_0_DOCS.ZIP binary
This zip file contains the following documents in both .docx and .pdf
VLE Micro_Lab_Micro_Interface_Release_1.0_Technical_Manual
VLE Micro_Lab_Micro_Interface_Release_1.0_User_Guide
VLE Micro_Deployment_Installation_Roll Back_Back Out_Guide
elements functions.
The VistA Documentation Library (VDL) web site will also contain the
above referenced documents. This website is usually updated within 1-3
days of the patch release date. The VDL web address for Laboratory
Universal Interface user documentation is:
http://www.domain.ext/vdl/application.asp?appid=120
Pre/Post Installation Overview
------------------------------
There are no Pre/Post installation routines associated with this
patch.
Pre-Installation Instructions
-----------------------------
This patch may be installed with users on the system although it is
recommended that it be installed during non-peak hours to minimize
potential disruption to users. This patch should take less than one
minute to install.
NOTE: Kernel patches must be current on the target system to avoid
RISK ANALYSIS: Changes made by patch LR*5.2*480 have no effect on Blood
problems loading and/or installing this patch.
==================== Installation Instructions ====================
1. Use the 'INSTALL/CHECK MESSAGE' option of the PackMan menu. This
option will load the KIDS patch onto your system.
2. The patch has now been loaded into a Transport global on your
Bank software functionality, therefore RISK is none.
system. You now need to use KIDS to install the transport global.
3. On the 'Kernel Installation & Distribution System' Menu (KIDS),
select the 'Installation' menu.
4. Use the 'Verify Checksum in Transport Global' option and verify
that all routines have the correct checksums.
5. On the KIDS menu, under the 'Installation' menu, use the following
options:
Print Transport Global
Compare Transport Global to Current System
Backup a Transport Global
If you wish to preserve a copy of the routines exported in this
patch prior to installation, you should use the 'Backup a
Transport Global' option at this time. You may also compare the
routines in your production account to the routines in the patch
by using the 'Compare a Transport Global to Current System'
option.
VALIDATION REQUIREMENTS BY OPTION: Because of the nature of the changes
6. Use the 'Install Package(s)' option under the 'Installation' menu
and select the package 'LR*5.2*480.
When Prompted "Want KIDS to INHIBIT LOGONs during the install?
NO//", respond NO.
When Prompted "Want to DISABLE Scheduled Options, Menu Options,
and Protocols? NO//", respond NO.
made, no specific validation requirements exist as a result of
Post-Installation Instructions
------------------------------
There are no post-installation tasks.
installation of this patch.
- Microbiology accession is being removed from the 'Incomplete
Remedy Ticket Overview:
==========================
N/A
New Service Requests (NSR)s):
==================
N/A
Participating Test Sites:
=========================
test status report (LRWRKINC)' option when a PRELIMINARY result is
James A. Haley Veterans Hospital
Iowa City VA Healthcare System
Lexington VA Medical Center
Problem:
--------
When releasing a microbiology preliminary report using the new
Universal Interface release options the status of the accession was being
set to complete and was then no longer available on the Incomplete Test
Status list report.
released when using the Universal Interface Microbiology release options.
Resolution:
-----------
The patch modifies LRVR0 at ACCEPT+68 and adds the following three lines
of code:
;
; If Lab UI interface then prompt user for accession test
complete date /time in EC3^LRMIEDZ2 call.
I LRINTYPE=1 S LRFIFO=0
Routine: LRVR0
Problem:
--------
When a retransmission of microbiology results are received the accepting
of the retransmission values was causing the existing
organism, comments and susceptibilities to be deleted. This is expected
functionality when handling verified microbiology reports
from external reference laboratories.
However when handling result transmission from a Universal Interface
microbiology instrument the user before the report is
verified can enter additional information that needs to be preserved.
Resolution:
-----------
The patch modifies the handling of existing results based on the type of
interface. LEDI types of interfaces will continue to
remove previous information prior to storing new information from the
Blood Bank Review:
reference laboratory. Universal Interfaces type interfaces
will maintain existing information entered by Laboratory or received via
prior transmissions and merge the new transmission with
the existing data before presenting to lab user for review.
Routine: LRVRMI2A modified labels N9A, N12A, N17
Routine: LRVRMI4A modified labels N3A, N6A, N6B
==================
Software and Documentation Retrieval Instructions:
----------------------------------------------------
Software being released as a Packman message and/or documentation
describing the new functionality introduced by this patch are available.
The preferred method is to retrieve files from download.vista.domain.ext.
This transmits the files from the first available server. Sites may
also elect to retrieve files directly from a specific server.
Sites may retrieve the software and/or documentation directly using
EFFECT ON BLOOD BANK FUNCTIONAL REQUIREMENTS: Patch LR*5.2*480 does not
Secure File Transfer Protocol (SFTP) from the ANONYMOUS.SOFTWARE
directory at the following OI Field Offices:
Albany: fo-albany.domain.ext
Hines: fo-hines.domain.ext
Salt Lake City: fo-slc.domain.ext
Documentation can also be found on the VA Software Documentation Library
at: http://www4.domain.ext/vdl/
|