Name | Value |
---|---|
NAME | COULTER +5 |
PROGRAM | LACOLT5 |
NEW DATA | D NEW^LASET |
RESTART | D RESTART^LASET |
LABORATORY TESTS |
|
ENTRY for LAGEN ROUTINE | Tray-cup or sequence x-ref |
CROSS LINKED BY | IDE |
INTERFACE NOTES | Coulter S+4,S+5,S+6 all may use the program, LACOLT5. value will be 1 less, and should be cross linked by ID. The ticket flag transmission mode should be set to OFF on the instrument. The COUTLER JT3 will run with either the LACOLT5 or the LACOLT6 routines. If you are not running the EO# and BA# tests you can use the LACOLT5. If you are running these tests you will need to use LACOLT6 as when these tests are activated the "," is not used between the test name and the result. Instruments using DT WITH DH INTERFACE and Coulter S+ Jr. (the format in ^LA will not have a ',' between the test name and the results) use LACOLT6. Parameters will depend on wether a patient id # is transmitted (this is controlled by a dip switch setting inside the Coulter's CRT). If a patient id is transfered, WBC will have for Param 1: S V=Y(5) and should be cross linked by IDE. If a patient id is not transfered, Param 1: S V=Y(4) ie. the subscript |