| DESCRIPTION OF ENHANCEMENTS |
This patch (VBEC*2*1) is being released in conjunction with the Vista
The VistA patch is released with a seven (7) day compliance date.
=======================================================================
Veterans Health Administration (VHA) Customer Documents:
VistA Blood Establishment Computer Software (VBECS) 2.3.0 Release
Notes.docx
VistA Blood Establishment Computer Software (VBECS) 2.3.0 Release
Notes.pdf
VistA Blood Establishment Computer Software (VBECS) 2.3.0 Technical
Manual-Security Guide.docx
VistA Blood Establishment Computer Software (VBECS) 2.3.0 Technical
Manual-Security Guide.pdf
VistA Blood Establishment Computer Software (VBECS) 2.3.0 User Guide.docx
VistA Blood Establishment Computer Software (VBECS) 2.3.0 User Guide.pdf
VistA Blood Establishment Computer Software (VBECS) 2.3.0 Admin User
Guide.docx
VistA Blood Establishment Computer Software (VBECS) 2.3.0 Admin User
Guide.pdf
VBECS Known Defects and Anomalies.docx
VBECS Known Defects and Anomalies.pdf
BLOOD BANK CLEARANCE
PATCH INSTALLATION:
===================
Pre/Post installation Overview:
===============================
Refer to the VistA Blood Establishment Computer Software (VBECS) 2.3.0
Release Notes document for pre and post installation information and
instructions.
====================
Pre-Installation Instructions:
==============================
Create a test with the following information prior to patch installation.
NAME: VBEC PATIENT REFLEX TEST TYPE: NEITHER
SUBSCRIPT: WORKLOAD HIGHEST URGENCY ALLOWED: ROUTINE
PRINT NAME: VBEC PATIENT REFLEX CREATION DATE: <Use current date>
Installation Instructions:
==========================
EFFECT ON BLOOD BANK FUNCTIONAL REQUIREMENTS: Patch VBEC*2*1 contains
This patch may be loaded with users on the system.
Installation will take less than 5 minutes.
1) Use the INSTALL/CHECK MESSAGE option on the PackMan menu to unload
the KIDS distribution included with this message.
2) From the Kernel Installation and Distribution System Menu,
select the Installation menu.
3) From this menu, you may elect to use the following options
(when prompted for the INSTALL NAME, enter VBEC*2*1).
a) Backup a Transport Global - this option will create a backup
changes to a package referenced in VHA OI SEPG SOP 192-023 Review of
message of any routines exported with the patch. It will NOT
backup any other changes such as DDs or templates.
b) Compare Transport Global to Current System - this option will
allow you to view all changes that will be made when the patch
is installed. It compares all components of the patch
(routines, DDs, templates, etc.).
c) Verify Checksums in Transport Global - this option will allow
you to ensure the integrity of the routines that are in the
transport global.
d) Print Transport Global - this option will allow you to view
VISTA Patches for Effects on VISTA Blood Bank Software. This patch does
the components of the KIDS build.
4) Use the Install Package(s) option and select the package
VBEC*2*1.
5) If prompted 'This Patch is for Version 2, you are running
Version 1 Want to continue installing this build? No//' respond
YES.
6) When prompted 'Want KIDS to INHIBIT LOGONs during the install?
NO//' respond NO.
7) When prompted 'Want to DISABLE Scheduled Options, Menu Options,
and Protocols? NO//' respond NO.
not alter or modify any VistA Blood Bank software design safeguards or
safety critical elements functions.
Blood Establishment Computer Software (VBECS) VBECS 2.3.0 server patch.
RISK ANALYSIS: Changes made by patch VBEC*2*1 have no adverse effect on
VistA Blood Bank software functionality, therefore RISK is none.
VALIDATION REQUIREMENTS:
========================
The VBECS 2.3.0 update and VBEC*2*1 require customer validation in the
local Test account prior to installation in Production to meet Food &
Drug Administration (FDA) requirements. The user validates the Blood Bank
software for its intended use prior to its acceptance and use for patient
care. The VBECS 2.3.0 Release Notes contain descriptions of the changes
made in this patch and includes validation scenarios.
It is imperative that the installation of VBEC*2*1 occur prior to the
installation of the VBECS 2.3.0 server patch. Coordinate the installation
with Pathology & Laboratory Medicine Service staff to allow adequate time
for impact analysis and testing the changes prior to installation in
Production.
Files & Fields Associated:
==========================
Installation of this patch will have no effect on sites who have not
N/A
FORMS ASSOCIATED:
=================
N/A
MAIL GROUPS ASSOCIATED:
=======================
N/A
installed VBECS.
OPTIONS ASSOCIATED:
===================
N/A
PROTOCOLS ASSOCIATED:
=====================
N/A
SECURITY KEYS ASSOCIATED:
=========================
N/A
TEMPLATES ASSOCIATED:
=====================
N/A
ADDITIONAL INFORMATION:
=======================
N/A
A full description of the VistA patch changes and implementation process
NEW SERVICE REQUESTS (NSRS):
============================
N/A
PATIENT SAFETY ISSUES (PSIS):
=============================
N/A
DEFECT TRACKING SYSTEM TICKET(S) & OVERVIEW:
============================================
are included in the VBECS 2.3.0 Release Notes.
I5868220FY15
I7744342FY16
I6114955FY16
I7083226FY16
I17762673FY18
TEST SITES:
===========
MILWAUKEE, WI
PHILADELPHIA, PA
SAN FRANCISCO, CA
HINES, IL
IOWA CITY, IA
DOCUMENTATION RETRIEVAL INSTRUCTIONS:
=====================================
Documentation describing the new functionality introduced by this patch
is available. Documentation can be found on the VA Software Documentation
Library at: https://www.domain.ext/vdl/application.asp?appid=182
|