| DESCRIPTION OF ENHANCEMENTS |
This patch to the Laboratory package does not contain any changes to the
Solution: The code has been validated, brought up to national
============================
VAMC's are encouraged to use their FTP capability to obtain the
Documentation PDF files from one of the following OI Field Office's
ANONYMOUS directories.
OI FIELD OFFICE FTP ADDRESS DIRECTORY
=============== =========== ==========
HINES 152.129.1.110 [ANONYMOUS.SOFTWARE]
SALT LAKE CITY 152.131.2.1 [ANONYMOUS.SOFTWARE]
ALBANY 152.127.1.5 [ANONYMOUS.SOFTWARE]
standard and released as a Class I national patch.
File names are:
BINARY FILES
LA7DIG.PDF -- CareVue Installation Guide
LA7DIS.PDF -- CareVue Interface Specifications
LA7DTP.PDF -- CareVue Interface Test Plan
The CareVue interface utilizes VISTA'S Laboratory Test Result system to
send Observation Request Unsolicited (ORU) messages to the CareVue system.
The interface messages are based on ward location and passed to an
interface engine. When Lab results are verified and the requesting ward
is identified as having a supported entry in the new CareVue Foreign
Interface file #62.487, the lab results are forwarded, via HL7, to the
interface engine and then on to the CareVue Clinical Information
Management System.
Each facility will need to refer to the CareVue Interface Specifications
manual LA7DIS.PDF as noted above for further requirements regarding
interface engines.
This patch REQUIRES that patch LR*5.2*254 be installed in order for the
INSTALLATION INSTRUCTIONS
==========================
The patch is distributed in KIDS format in a PackMan message.
It is recommended that this patch be installed in a test environment first
before installing into production. Each environment (test and production)
uses a separate database and a separate integration interface. Because of
this the data that is communicated from VistA may not always be consistent
with the data that will be communicated from the production environment.
Another impact is that patient data may have to be manually entered into
CareVue interface to work properly.
the VistA test environment in order to test the proper handling of
messages going to VistA.
The install time for this patch is less than 2 minutes. This patch can be
installed when Laboratory users are on the system.
1. All routines installed with this patch are new to the system and
mapping is not recommended.
2. Use the 'INSTALL/CHECK MESSAGE' option of the PackMan menu.
This option will load the KIDS patch onto your system.
3. The patch has now been loaded into a Transport global on your
system. You now need to use KIDS to install the transport
global.
On the KIDS menu, under the 'Installation' menu, use the
following options:
Print Transport Global
Compare Transport Global to Current System
An interface engine must exist to connect CareVue (or other clinical
Verify Checksums in Transport Global
Backup a Transport Global
4. Use the 'Install Package(s)' option under the 'Installation' men
and select the package 'LA*5.2*53'. When prompted 'Want to
DISABLE Scheduled Options, Menu Options, and Protocols? YES//',
choose 'NO'.
5. At the DEVICE prompt, you can send your output to any printer of
HFS file, display it on the terminal screen, or queue it to run
system) and Vista using TCP/IP and HL7 messages. Three possible contacts
at a scheduled time.
6. No options need to be placed out of service.
7. This installation automatically adds the two new protocols to
the PROTOCOL file (#101). They are: LA7D CARELIFE RESULTS
LA7D CARELIFE SERVER
8. User will need to enter your facility name in the HL7
APPPLICATION PARAMETER file (#771).
for this interface engine are STC, Inc. MSI Solutions Inc. and Agilent
NOTE: Fields marked with <-- below need to have information entered.
Select OPTION: ENTER OR EDIT FILE ENTRIES <--
INPUT TO WHAT FILE: FOREIGN INTERFACE// HL7 APPLICATION PARAMETER <--
(34 entries)
EDIT WHICH FIELD: ALL// FACILITY NAME <--
THEN EDIT FIELD: <RET>
Select HL7 APPLICATION PARAMETER NAME: DHCP CARELIFE ACTIVE <--
FACILITY NAME: 428// 428 enter your facility name here <--
Technologies.
Select HL7 APPLICATION PARAMETER NAME: MRTS CARELIFE ACTIVE <--
FACILITY NAME: 642// 428 enter your facility name here <--
Select HL7 APPLICATION PARAMETER NAME: <RET>
9. The new mail group CARELIFE RESULT ERROR was created as part of
the installation. However the entry in the MAIL GROUP file (3.8)
is only a stub and requires the site to perform the following edit:
NOTE: Fields marked with <-- below need to have information entered.
VISTA BLOOD BANK Software as defined by VHA DIRECTIVE 99-053 titled VISTA
Select MAIL GROUP NAME: CARELIFE RESULT ERROR <--
NAME: CARELIFE RESULT ERROR Replace
Select MEMBER: CROSS,PAM <--
Are you adding 'CROSS,PAM' as a new MEMBER (the 1ST for this MAIL
GROUP)
? No// Y (Yes) <--
TYPE:
Select MEMBER: <RET>
DESCRIPTION:
No existing text
STC, Inc. can be contacted at:
Edit? NO// YES <--
All error report for CAREVUE ORU Messages <--
TYPE: public// PUBLIC <--
ORGANIZER: CROSS,PAM <--
COORDINATOR: CROSS,PAM <--
Select AUTHORIZED SENDER:
ALLOW SELF ENROLLMENT?: NO// NO <--
REFERENCE COUNT: <RET>
LAST REFERENCED: <RET>
Richard Garcia
RESTRICTIONS: INDIV/LOCAL/ORGANIZER// <RET>
Select MEMBER GROUP NAME: <RET>
Select REMOTE MEMBER: <RET>
Select DISTRIBUTION LIST: <RET>
Select FAX RECIPIENT: <RET>
Select FAX GROUP: <RET>
Select MAIL GROUP NAME: <RET>
10. The FOREIGN INTERFACE file (#62.487) was put on the system during
voice: 626.471.6000 x6755
the installation process but it does not contain any data. You
must enter your site's data concerning your CareVue wards. This
option will need to be added to users secondary menus under the
namespace LA7D.
NOTE: Fields marked with <-- below need to have information entered.
Select MailMan Menu Option: Edit Foreign Interface Parameters
Enter BEDSIDE MONITOR INTERFACE NAME: CAREVUE <--
Enter CAREVUE WARD: ICU <-- enter your CareVue wards here
fax: 626.471.6104
Enter CAREVUE WARD: ADMISSIONS <--
Enter CAREVUE WARD: ^
**NOTE**
All wards in the hospital location file (#44) that will be
used for the CareVue interface must have a UNIQUE abbreviation for
each ward (Inactive and Active)
11. Set up Logical Link for HL7 parameters. This will need to contain
the outgoing device name used from VistA to the interface engine.
email: rmg@stc.com
Sites should ensure that entries corresponding to their unique
system requirements be input at the recommended places.
NOTE: Fields marked with <-- below need to have information entered.
Select Systems Manager Menu Option: HL7 Main Menu
Select HL7 Main Menu Option: V1.6 OPTIONS
Slect V1.6 OPTIONS Option: Interface Workbench
At the bottom of the screen you will see:
Select Tool: Jump to Next App// SL Show Logical Links
Select Tool: Jump to Next Link// CL Create Logical Link
Select HL LOGICAL LINK NODE: VOICE1 <--
Logical Link information
------------------------
Queue Size: 245 <--
MSI Solutions, Inc. can be contacted at:
LLP Parameter: HLLP <--
LLP Parameters
--------------
LLP Type: HLLP <--
Device: OG_LAB_PROD <-- this will be your device name
Version ID: 2.3 <--
Block Size: 245 <--
Read Timeout: 30 <--
ACK Timeout: 2 <--
Bob Kelliher
Re-transmission Attempts: 3 <--
Start Block Character: <RET>
End Block Character: <RET>
Select Tool: Jump to Next Link// ^
NOTE: Add the Logical Link just created to the LA7D CARELIFE
RESULTS protocol.
NOTE: Fields marked with <-- below need to have information entered.
voice: 678.256.4505
Select OPTION: ENTER OR EDIT FILE ENTRIES
INPUT TO WHAT FILE: PROTOCOL// PROTOCOL <-- (3569 entries)
EDIT WHICH FIELD: ALL// NAME <--
THEN EDIT FIELD: LOGICAL LINK <--
THEN EDIT FIELD:
Select PROTOCOL NAME: LA7D CARELIFE RESULTS <--
This client is tied to the MRTS CARELIFE entry
BLOOD BANK SOFTWARE.
fax: 678.256.4677
NAME: LA7D CARELIFE RESULTS Replace <RET>
LOGICAL LINK: OG_LAB_PROD <-- Enter your Logical Link that was set up
12. Follow the Interface Specifications Manual LA7DIS.PDF for
interface engine requirements.
13. Follow the CareVue Test Plan Manual LA7DTP.PDF to test the CareVue
package.
email: bob.kelliher@msi-solutions.com
Agilent Technologies can be contacted at:
Linda Vanderhulst
voice: 847.342.2521
email: linda_vanderhulst@agilent.com
About this software:
====================
1. An option to enter/delete supported wards for the FOREIGN
INTERFACE file (#62.487) is exported with this version of the patch.
**NOTE**
All wards in the hospital location file (#44) that will be
used for the CareVue interface must have a UNIQUE abbreviation for
each ward (Inactive and Active)
2. Resulting from this installation, the entries DHCP CARELIFE and
MRTS CARELIFE will automatically be added to the HL7 APPLICATION
This patch will install the CareVue interface package under the Namespace
PARAMETER file (#771). These entries will be used to send HL7
messages to the CareVue system. These parameters also need to be
edited to enter your facility name.
3. The following file is included in this release. It is in the LA7D
namespace, also listed in the main installation procedure:
File #62.487 File name FOREIGN INTERFACE
4. The following mail group will be automatically added to the MAIL
LA7D as a Class I national patch.
GROUP file (#3.8) during this installation: CARELIFE RESULT ERROR
5. This installation automatically adds the two new protocols to the
PROTOCOL file (#101). They are: LA7D CARELIFE RESULTS
LA7D CARELIFE SERVER
6. Sites will need to set up a Logical Link for HL7 parameters.
7. The routines exported with this release, listed in the Routines
section of this document, are customized to send verified
Laboratory data to CareVue for all patients located on a CareVue
ward.
Associated NOIS
===============
None
Test Sites
===============
1. Problem: Current CareVue interface is Class III software.
Philadelphia, PA
Martinsburg, WVA
ROUTINE SUMMARY
=================
The following is a list of routines included in this patch. The second
line of each routine now looks like:
<tab>;;5.2;LAB SERVICE;**[patch list]**;Sep 27, 1994
CHECK^XTSUMBLD Results
Routine Name Before Patch After Patch Patch List
------------ ------------ ----------- ------------
LA7DVC NEW 7338732 53
LA7DVEXT NEW 5483063 53
LA7DVM NEW 6983644 53
LA7DEDT NEW 3756680 53
CareVue installation manuals
|